Surface Plasmon Resonance
How a 50 nm gold film turns a refractive-index change of 10−6 into a measurable shift in a reflectivity dip, and how that shift becomes a binding constant. Every plot on this page is computed live from the equations shown.
A surface plasmon is a bound TM wave
A surface plasmon polariton (SPP) is a collective oscillation of conduction electrons coupled to an electromagnetic field, confined to the interface between a metal (\(\varepsilon_m\)) and a dielectric (\(\varepsilon_d\)). Solving Maxwell's equations for a wave that propagates along \(x\) and decays in \(\pm z\), the continuity of \(H_y\) and \(E_x\) leaves a single non-trivial condition, and only for TM (p) polarisation. TE modes do not exist on a single flat interface.
With \(\varepsilon_m = \varepsilon_m' + i\varepsilon_m''\), the mode has a finite propagation length and decays exponentially away from the interface. The decay into the dielectric sets the sensing depth: SPR only sees what lies within roughly one \(\delta_d\) of the gold.
Try 633 nm vs 850 nm: longer wavelengths give sharper resonances (larger |ε′|/ε″) and a longer propagation length, but a deeper field.
Coupling light: closing the momentum gap
Because \(\varepsilon_m<0\) with \(|\varepsilon_m|>\varepsilon_d\), the square root in the dispersion relation exceeds \(\sqrt{\varepsilon_d}\): the SPP always carries more in-plane momentum than a photon of the same frequency in the dielectric. Light incident from the dielectric side can never excite it. The dispersion curve lies entirely to the right of the light line.
Kretschmann–Raether
Metal film deposited on the prism. Light tunnels through ~50 nm Au. The standard geometry for commercial biosensors.
Otto
A dielectric gap of ~λ separates prism and metal. Useful for bulk metals or when the metal surface must stay untouched.
Grating / waveguide
Diffraction or a guided mode supplies the extra momentum. Compatible with planar chips and fibres.
The gold line is the in-plane wavevector the prism supplies. Where it crosses the SPP curve, a plasmon is excited at that frequency. It can never cross while \(n_p\sin\theta < n_d\).
The reflectivity dip, computed exactly
For the Kretschmann stack (prism | Au | optional adlayer | bulk), the p-polarised reflectance follows from the Fresnel coefficients for each interface, combined recursively from the bottom layer up. The dip is where energy leaks into the SPP and is dissipated in the gold. It is deepest when radiative loss (set by film thickness) equals internal absorption loss: critical coupling.
Things to try: sweep d from 30 to 70 nm and watch the dip pass through critical coupling (Rmin → 0 near 47–50 nm at 633 nm). Switch to SF10: the resonance moves to smaller angles and the angular sensitivity drops. Add a 3 nm protein layer: the shift is a fraction of the bulk shift you would get from the same Δn filling the whole field.
Sensitivity, resolution and what binds at the surface
A sensor's response is the product of how far the resonance moves per unit refractive index and how far the index moves per bound molecule. Keep these separate: instrument papers quote bulk sensitivity, but binding experiments are governed by the surface term.
| Quantity | Typical value (Au, 633–850 nm, water) | Comment |
|---|---|---|
| Angular sensitivity Sθ | ≈ 100–150 °/RIU (BK7) | Compute yours in the lab above |
| Spectral sensitivity Sλ | ≈ 2 000–10 000 nm/RIU | Rises steeply toward the NIR |
| Resolution (good instruments) | 10−6–10−7 RIU | Noise-limited, needs temperature control (dn/dT of water ≈ −1×10−4 K−1) |
| Response unit (Biacore convention) | 1 RU ≈ 10−6 RIU ≈ 10−4 ° | ≈ 1 pg mm−2 of protein |
| Field penetration δd | ≈ 150–400 nm | Molecules far from the chip contribute little |
Temperature is often the dominant noise source. A 0.01 K drift in water is ≈ 10−6 RIU, i.e. about 1 RU. That is why reference channels and thermostatting matter as much as optics.
From sensorgram to rate constants
In a biosensor experiment a ligand is immobilised on the chip and an analyte at concentration \(C\) flows over it. The response \(R(t)\) is proportional to bound mass. For a 1:1 interaction \(A + L \rightleftharpoons AL\) under pseudo-first-order conditions (constant \(C\)):
Rule of thumb for design: span concentrations from ~0.1× to ~10× KD, keep Rmax low (tens of RU for kinetics) to limit mass transport and rebinding, and inject long enough that the highest concentration approaches equilibrium. If kobs stops scaling linearly with C, suspect mass transport or heterogeneity before reaching for a two-state model.
Practice and pitfalls
Double referencing
Subtract a reference flow cell (same surface, no ligand) to remove bulk RI and non-specific binding, then subtract buffer-only (blank) injections to remove systematic drift.
Solvent matching
1% DMSO shifts the bulk signal by roughly 1 000 RU. Match DMSO exactly and run a solvent-correction curve for small-molecule work.
Mass transport
Symptoms: linear association phases, ka that depends on flow rate or ligand density. Remedy: lower density, higher flow, or fit with a two-compartment model.
Surface chemistry
Carboxymethyl dextran (EDC/NHS amine coupling), SAMs of alkanethiols, capture formats (His-tag/NTA, streptavidin–biotin, anti-Fc). Orientation and activity decide your effective Rmax.
Regeneration
Scout mild to harsh conditions (low-pH glycine, high salt, SDS). A good regeneration returns baseline without losing ligand activity over many cycles.
Fitting honestly
Fit the whole concentration series globally, inspect residuals, compare kd/ka with steady-state KD, and report χ², replicates and the model used.
Film deposition notes
Gold is usually evaporated or sputtered at 45–50 nm on a 1–2 nm Cr or Ti adhesion layer. The adhesion layer is absorbing and broadens the dip, so keep it thin or model it explicitly (add a fourth layer to the Fresnel recursion). Silver gives sharper resonances but oxidises; bimetallic Ag/Au films are a common compromise.
Beyond the flat-film sensor
| Technique | Principle | Why use it |
|---|---|---|
| SPR imaging (SPRi) | Fixed θ, image reflectance with a camera | Hundreds of spots in parallel; arrays, screening |
| Long-range SPR | Symmetric dielectric around a thin (~20 nm) metal film couples the two interface modes | Narrower dip, deeper field, higher FOM |
| Localised SPR (LSPR) | Particle resonance, Fröhlich condition \(\operatorname{Re}\varepsilon_m = -2\varepsilon_d\) for a sphere | No prism; decay length ~10 nm, very surface-selective |
| Fibre / waveguide SPR | Guided mode supplies the in-plane momentum | Miniaturisation, remote and in-vivo sensing |
| Phase-sensitive SPR | Track the steep phase jump of rp at resonance | Resolution down to 10−8 RIU |
| SPR fluorescence (SPFS) / SERS | Use the field enhancement at the surface | Detect small or dilute analytes beyond refractometric limits |
Self-check
0 / 6 answered
Reading list
- H. Raether, Surface Plasmons on Smooth and Rough Surfaces and on Gratings, Springer Tracts in Modern Physics 111 (1988). The reference for the physics.
- S. A. Maier, Plasmonics: Fundamentals and Applications, Springer (2007). Best first textbook.
- E. Kretschmann, H. Raether, Z. Naturforsch. A 23, 2135 (1968); A. Otto, Z. Phys. 216, 398 (1968). The two coupling geometries.
- P. B. Johnson, R. W. Christy, Phys. Rev. B 6, 4370 (1972). Optical constants of noble metals used here.
- B. Liedberg, C. Nylander, I. Lundström, Sens. Actuators 4, 299 (1983). First SPR biosensing demonstration.
- J. Homola, S. S. Yee, G. Gauglitz, Sens. Actuators B 54, 3 (1999); J. Homola, Chem. Rev. 108, 462 (2008). Comprehensive sensor reviews.
- L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, S. S. Yee, Langmuir 14, 5636 (1998). Quantitative adlayer interpretation.
- D. G. Myszka, J. Mol. Recognit. 12, 279 (1999). Improving biosensor analysis: practical guidelines.
- Interactive optical constants: refractiveindex.info.