Surface Plasmon Resonance Field Guide
Graduate primer · optics → biosensing

Surface Plasmon Resonance

How a 50 nm gold film turns a refractive-index change of 10−6 into a measurable shift in a reflectivity dip, and how that shift becomes a binding constant. Every plot on this page is computed live from the equations shown.

analyte, ε_d = n_s² evanescent field δ_d ≈ 100–300 nm Au, d ≈ 50 nm prism n_p TM light R_p(θ) θ Kretschmann–Raether
01

A surface plasmon is a bound TM wave

A surface plasmon polariton (SPP) is a collective oscillation of conduction electrons coupled to an electromagnetic field, confined to the interface between a metal (\(\varepsilon_m\)) and a dielectric (\(\varepsilon_d\)). Solving Maxwell's equations for a wave that propagates along \(x\) and decays in \(\pm z\), the continuity of \(H_y\) and \(E_x\) leaves a single non-trivial condition, and only for TM (p) polarisation. TE modes do not exist on a single flat interface.

Boundary condition (TM)
\[ \frac{k_{z,d}}{\varepsilon_d} + \frac{k_{z,m}}{\varepsilon_m} = 0, \qquad k_{z,i}^2 = \varepsilon_i\,k_0^2 - k_x^2,\quad k_0=\frac{\omega}{c}=\frac{2\pi}{\lambda} \]
Dispersion relation
\[ k_{sp} = k_0\sqrt{\frac{\varepsilon_m\,\varepsilon_d}{\varepsilon_m+\varepsilon_d}} \quad\Longrightarrow\quad \text{bound mode requires } \operatorname{Re}\varepsilon_m < -\varepsilon_d \]

With \(\varepsilon_m = \varepsilon_m' + i\varepsilon_m''\), the mode has a finite propagation length and decays exponentially away from the interface. The decay into the dielectric sets the sensing depth: SPR only sees what lies within roughly one \(\delta_d\) of the gold.

Propagation length (intensity, 1/e)
\[ L_{sp} = \frac{1}{2\,\operatorname{Im}k_{sp}} \]
Penetration depth (field, 1/e)
\[ \delta_i = \frac{1}{\left|\operatorname{Im}k_{z,i}\right|} = \frac{1}{k_0}\left|\operatorname{Re}\sqrt{\frac{\varepsilon_i^2}{\varepsilon_m+\varepsilon_d}}\right|^{-1} \]
Live · SPP at Au / dielectricAu optical constants interpolated from Johnson & Christy (1972)

Try 633 nm vs 850 nm: longer wavelengths give sharper resonances (larger |ε′|/ε″) and a longer propagation length, but a deeper field.

|E_z| in dielectric|E_z| in gold (×10 depth scale)
02

Coupling light: closing the momentum gap

Because \(\varepsilon_m<0\) with \(|\varepsilon_m|>\varepsilon_d\), the square root in the dispersion relation exceeds \(\sqrt{\varepsilon_d}\): the SPP always carries more in-plane momentum than a photon of the same frequency in the dielectric. Light incident from the dielectric side can never excite it. The dispersion curve lies entirely to the right of the light line.

Resonance (phase-matching) conditions
\[ \underbrace{n_p\,k_0\sin\theta}_{\text{prism (ATR)}} = \operatorname{Re}k_{sp} \qquad\qquad \underbrace{n_d\,k_0\sin\theta + m\frac{2\pi}{\Lambda}}_{\text{grating, order } m} = \operatorname{Re}k_{sp} \]

Kretschmann–Raether

Metal film deposited on the prism. Light tunnels through ~50 nm Au. The standard geometry for commercial biosensors.

Otto

A dielectric gap of ~λ separates prism and metal. Useful for bulk metals or when the metal surface must stay untouched.

Grating / waveguide

Diffraction or a guided mode supplies the extra momentum. Compatible with planar chips and fibres.

Live · dispersion diagram (lossless Drude metal)\(\varepsilon_m(\omega)=1-\omega_p^2/\omega^2\)

The gold line is the in-plane wavevector the prism supplies. Where it crosses the SPP curve, a plasmon is excited at that frequency. It can never cross while \(n_p\sin\theta < n_d\).

SPP dispersionlight line in dielectricprism: k = n_p sinθ ω/c
03

The reflectivity dip, computed exactly

For the Kretschmann stack (prism | Au | optional adlayer | bulk), the p-polarised reflectance follows from the Fresnel coefficients for each interface, combined recursively from the bottom layer up. The dip is where energy leaks into the SPP and is dissipated in the gold. It is deepest when radiative loss (set by film thickness) equals internal absorption loss: critical coupling.

Interface and film coefficients (p-polarisation)
\[ r_{ij} = \frac{k_{z,i}/\varepsilon_i - k_{z,j}/\varepsilon_j}{k_{z,i}/\varepsilon_i + k_{z,j}/\varepsilon_j}, \qquad r_{012} = \frac{r_{01} + r_{12}\,e^{2ik_{z,1}d}}{1 + r_{01}\,r_{12}\,e^{2ik_{z,1}d}}, \qquad R_p = |r_{012}|^2 \] \[ k_{z,j} = k_0\sqrt{\varepsilon_j - n_p^2\sin^2\theta}\quad(\operatorname{Im}k_{z,j}\ge 0) \]
Lorentzian approximation near resonance (Raether)
\[ R_p(k_x) \approx 1 - \frac{4\,\Gamma_i\,\Gamma_{rad}}{\left(k_x - \operatorname{Re}k_{sp}-\Delta k\right)^2 + \left(\Gamma_i+\Gamma_{rad}\right)^2},\qquad \Gamma_i=\operatorname{Im}k_{sp},\ \Gamma_{rad}\propto e^{-2|k_{z,1}|d} \]
Live · Kretschmann reflectivity lab
current stackreferencecritical angle

Things to try: sweep d from 30 to 70 nm and watch the dip pass through critical coupling (Rmin → 0 near 47–50 nm at 633 nm). Switch to SF10: the resonance moves to smaller angles and the angular sensitivity drops. Add a 3 nm protein layer: the shift is a fraction of the bulk shift you would get from the same Δn filling the whole field.

04

Sensitivity, resolution and what binds at the surface

A sensor's response is the product of how far the resonance moves per unit refractive index and how far the index moves per bound molecule. Keep these separate: instrument papers quote bulk sensitivity, but binding experiments are governed by the surface term.

Bulk sensitivity (angular)
\[ S_\theta = \frac{\partial\theta_{res}}{\partial n_s} \approx \frac{\varepsilon_m'\sqrt{-\varepsilon_m'}}{(\varepsilon_m'+n_s^2)\sqrt{\varepsilon_m'(n_s^2-n_p^2)-n_s^2n_p^2}} \]
Figure of merit and detection limit
\[ \mathrm{FOM} = \frac{S}{\mathrm{FWHM}},\qquad \mathrm{LOD}_n = \frac{3\,\sigma_{noise}}{S} \]
Thin adlayer response (Jung et al., 1998)
\[ \Delta\theta \approx S_\theta\,(n_a - n_s)\left[1 - e^{-2d_a/l_d}\right],\qquad l_d\approx \delta_d \]
Surface coverage (de Feijter)
\[ \Gamma = \frac{d_a\,(n_a-n_s)}{\mathrm{d}n/\mathrm{d}c},\qquad \mathrm{d}n/\mathrm{d}c \approx 0.18\text{–}0.19\ \mathrm{mL\,g^{-1}}\ \text{(proteins)} \]
QuantityTypical value (Au, 633–850 nm, water)Comment
Angular sensitivity Sθ≈ 100–150 °/RIU (BK7)Compute yours in the lab above
Spectral sensitivity Sλ≈ 2 000–10 000 nm/RIURises steeply toward the NIR
Resolution (good instruments)10−6–10−7 RIUNoise-limited, needs temperature control (dn/dT of water ≈ −1×10−4 K−1)
Response unit (Biacore convention)1 RU ≈ 10−6 RIU ≈ 10−4 °≈ 1 pg mm−2 of protein
Field penetration δd≈ 150–400 nmMolecules far from the chip contribute little

Temperature is often the dominant noise source. A 0.01 K drift in water is ≈ 10−6 RIU, i.e. about 1 RU. That is why reference channels and thermostatting matter as much as optics.

05

From sensorgram to rate constants

In a biosensor experiment a ligand is immobilised on the chip and an analyte at concentration \(C\) flows over it. The response \(R(t)\) is proportional to bound mass. For a 1:1 interaction \(A + L \rightleftharpoons AL\) under pseudo-first-order conditions (constant \(C\)):

Langmuir 1:1 model
\[ \frac{\mathrm{d}R}{\mathrm{d}t} = k_a\,C\,(R_{max}-R) - k_d\,R \] \[ \text{association: } R(t) = R_{eq}\left(1-e^{-k_{obs}t}\right),\quad k_{obs}=k_aC+k_d,\quad R_{eq}=\frac{R_{max}\,C}{C+K_D} \] \[ \text{dissociation: } R(t) = R_0\,e^{-k_d(t-t_0)},\qquad K_D=\frac{k_d}{k_a},\qquad t_{1/2}=\frac{\ln 2}{k_d} \]
Expected capacity and mass transport
\[ R_{max} = R_L\,\frac{M_A}{M_L}\,n_{sites},\qquad \frac{\mathrm{d}R}{\mathrm{d}t}=\frac{k_a C (R_{max}-R)-k_d R}{1+k_a(R_{max}-R)/k_m} \]
Live · sensorgram simulator5 concentrations, 2-fold series · injection 0–180 s · buffer to 600 s

Rule of thumb for design: span concentrations from ~0.1× to ~10× KD, keep Rmax low (tens of RU for kinetics) to limit mass transport and rebinding, and inject long enough that the highest concentration approaches equilibrium. If kobs stops scaling linearly with C, suspect mass transport or heterogeneity before reaching for a two-state model.

06

Practice and pitfalls

Double referencing

Subtract a reference flow cell (same surface, no ligand) to remove bulk RI and non-specific binding, then subtract buffer-only (blank) injections to remove systematic drift.

Solvent matching

1% DMSO shifts the bulk signal by roughly 1 000 RU. Match DMSO exactly and run a solvent-correction curve for small-molecule work.

Mass transport

Symptoms: linear association phases, ka that depends on flow rate or ligand density. Remedy: lower density, higher flow, or fit with a two-compartment model.

Surface chemistry

Carboxymethyl dextran (EDC/NHS amine coupling), SAMs of alkanethiols, capture formats (His-tag/NTA, streptavidin–biotin, anti-Fc). Orientation and activity decide your effective Rmax.

Regeneration

Scout mild to harsh conditions (low-pH glycine, high salt, SDS). A good regeneration returns baseline without losing ligand activity over many cycles.

Fitting honestly

Fit the whole concentration series globally, inspect residuals, compare kd/ka with steady-state KD, and report χ², replicates and the model used.

Film deposition notes

Gold is usually evaporated or sputtered at 45–50 nm on a 1–2 nm Cr or Ti adhesion layer. The adhesion layer is absorbing and broadens the dip, so keep it thin or model it explicitly (add a fourth layer to the Fresnel recursion). Silver gives sharper resonances but oxidises; bimetallic Ag/Au films are a common compromise.

07

Beyond the flat-film sensor

TechniquePrincipleWhy use it
SPR imaging (SPRi)Fixed θ, image reflectance with a cameraHundreds of spots in parallel; arrays, screening
Long-range SPRSymmetric dielectric around a thin (~20 nm) metal film couples the two interface modesNarrower dip, deeper field, higher FOM
Localised SPR (LSPR)Particle resonance, Fröhlich condition \(\operatorname{Re}\varepsilon_m = -2\varepsilon_d\) for a sphereNo prism; decay length ~10 nm, very surface-selective
Fibre / waveguide SPRGuided mode supplies the in-plane momentumMiniaturisation, remote and in-vivo sensing
Phase-sensitive SPRTrack the steep phase jump of rp at resonanceResolution down to 10−8 RIU
SPR fluorescence (SPFS) / SERSUse the field enhancement at the surfaceDetect small or dilute analytes beyond refractometric limits
08

Self-check

0 / 6 answered

09

Reading list

  1. H. Raether, Surface Plasmons on Smooth and Rough Surfaces and on Gratings, Springer Tracts in Modern Physics 111 (1988). The reference for the physics.
  2. S. A. Maier, Plasmonics: Fundamentals and Applications, Springer (2007). Best first textbook.
  3. E. Kretschmann, H. Raether, Z. Naturforsch. A 23, 2135 (1968); A. Otto, Z. Phys. 216, 398 (1968). The two coupling geometries.
  4. P. B. Johnson, R. W. Christy, Phys. Rev. B 6, 4370 (1972). Optical constants of noble metals used here.
  5. B. Liedberg, C. Nylander, I. Lundström, Sens. Actuators 4, 299 (1983). First SPR biosensing demonstration.
  6. J. Homola, S. S. Yee, G. Gauglitz, Sens. Actuators B 54, 3 (1999); J. Homola, Chem. Rev. 108, 462 (2008). Comprehensive sensor reviews.
  7. L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, S. S. Yee, Langmuir 14, 5636 (1998). Quantitative adlayer interpretation.
  8. D. G. Myszka, J. Mol. Recognit. 12, 279 (1999). Improving biosensor analysis: practical guidelines.